Objectives: To experimentally compare two fractional ablative CO2 laser handpieces intended for the treatment of large area burn scars.
Each handpiece coverage rate, depth of penetration and application time were measured and compared in a simulation model of large area burns scars using a dynamic/roller handpiece (small footprint) and a stationary/stamping handpiece (large footprint).
A 30 W fractional ablative CO2 laser was applied using 2 different handpieces and footprints on a A4 size paper stack.
The handpieces were a stationary (stamping) handpiece with 7 × 7 (49 pixels/square shape) and dynamic (roller) handpiece with 7 × 1 (7 pixels/single row shape).
For both handpieces the laser settings were fixed at "High" power (30 W), providing an energy level of 100 mJ/pixel. Both handpieces were applied perpendicular to the surface, with the process repeated for the dynamic handpiece with an angled operation.
The depth of laser penetration was assessed by the number of pages of paper having visible holes and burn area coverage time measured under each handpiece/condition.